| Display |
4-digit
LCD |
Range,
DO mg/l (ppm)
DO % Sat.
gaseous % Vol |
0-5.00
/ 0-10.00 / 0-20.0 / 0-100
0-50 / 0-100 / 0-200 / 0-1000
0-25.0 / 0-100
User can switch from one range to another within the units
selected by the appropriate jumper settings. |
| Signal
Input |
0-20
/ 4-20 mA. Max. load 750 ohm, user selectable. Galvanically
isolated. |
| Relay
Outputs |
- Low alarm, high alarm and control output.
- Low and high alarms have fixed hysteresis
of ±1%.
- Control output has adjustable hysteresis
from ±1% to ±10%.
- Single pole double throw, potential
free contacts.
- Maximum loading - 200 VA, 250 v AC.
|
| Power
Supply |
220/110/24V
AC, 50/60Hz. 1ph. or 24V DC. |
| Power
Consumption |
5VA |
| Dimensions |
19cm
(7 ") H x 1 cm (5 ") W x 9cm (3 ") D |
| Weight |
Approx.
1 kg (2.2 lb) |
| Enclosure |
IP66,
(NEMA4 equivalent). Robust, all-weather resistant plastic
enclosure. |
| Operating
Conditions |
-5°
to +40°C (23° to 104°F).
-40° to +40°C (-40° to 104°F) with optional
heater installed. |
| Mounting
Options |
Drip
shield with hand rail attachment |